Cleaning and maintenance

It is important to unplug the device before performing any cleaning.
Excess moisture may impair the proper operation of the device.
1. The sponge mesh filter:
The sponge mesh filter should be cleaned weekly to keep the oxygen concentrator running smoothly, or more
frequently if the environmental conditions requires it.
In order to clean the sponge mesh filter, please follow the following instructions :
a.
Open the air intake grid located on the back part of the
concentrator. Remove the sponge mesh filter and shake it to let
the dust fall.
b.
Rinse the sponge mesh filter thoroughly using clean water, remove
excess water, then allow to air dry completely.
c.
The dry mesh filter may be replaced and the air intake grid
reinstalled.
Having a second sponge mesh filter allows a quick
replacement while the other filter is drying. Always use
original spare parts. If the sponge filter needs to be replaced,
please contact your local distributor.
2. Compressor filter replacement:
The compressor filter is used to filter dust entering the compressor. It needs to be replaced regularly. It is
recommended to replace the filter every 4000 hours. The compressor filter needs to be purchased from a
SCALEO Medical distributor.
In order to replace the compressor filter, please follow the following instructions :
a)
The compressor filter cover is in the back of the device (see figure 4-1 below). Open the compressor
18
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5