Classification

IEC class
Degree of protection to concentrator components
against electrical shock
Degree of protection to concentrator components
against ingress of water while used
Mode of operation
Degree of safety for application in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Expected service life of the product
When the AC power supply voltage exceeds -15% to +10% of the rated voltage, the voltage is too
high, which may result in damage to the equipment. If the voltage is too low, the equipment may
not start. If the grid is unstable, please install a voltage regulator before use.
While in continuous operation, when the rated flow rate is reached, the time for reaching the specified
oxygen content should not exceed 30 minutes. 


Operating environment

Operating
temperature
Operating
relative humidity
Operating
atmospheric pressure
Surrounding environment
Recommended altitude for use
It is not recommended to use this concentrator in an environment above 3000 meters above sea
level, which may result in a decrease in oxygen concentration.
The ambient temperature of the oxygen machine is from 10°C to 37°C. Below 10°C, the compressor may
have difficulties starting. Above 37°C, the compressor may overheat and shorten the service life of the
compressor. If the device is stored in a very cold or very hot environment, it should be placed in the room
intended for use for 5 hours before use.
22
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5
Class II
Type BF
IP21
Continuous mode
Not suitable for such application
5 years
10 to
37°C

20% RH - 65% RH
80 kPa ~ 101 kPa
Dry and ventilated, free from dust and
any corrosive gases, without strong
electromagnetic interference
Below 3000 meters