IEC 60601
Immunity test
Test level
<5% U
Voltage dips,
(>95% dip in U
short
cycle 40% U
interruptions and
(60% dip in U
voltage variations
5 cycles 70% U
on power supply
(30% dip in U
input lines
25 cycles <5% U
IEC 61000-4-11
(>95% dip in U
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
U
is the A.C. mains voltage prior to application of the test level.
T
Conducted RF
3 Vrm
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
10V/m
IEC 61000-4-3
80 MHz to 2.7 GHz
26
Compliance level
<5% U
T
) for 0.5
(>95% dip in U
T
cycle 40% U
T
) for
(60% dip in U
T
5 cycles 70% U
T
) for
(30% dip in U
T
25 cycles <5% U
T
) for
(>95% dip in U
T
5 sec
5 sec
30 A/m
30 A/m
3 Vrm
10V/m
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5
Electromagnetic environment-guidance
T
) for 0.5
T
Mains power quality should be that of a typical home or
T
) for
hospital environment. If the user of the device requires
T
continued operation during power mains interruptions, it is
T
) for
recommended that the device be powered from an
T
uninterrupted power supply or a battery.
T
) for
T
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital or
Portable and mobile RF communications equipment should
be used no closer than the recommended 40 cm separation
distance to any part of the device, including cables.
Interference may occur in the vicinity of equipment marked
with the following symbol:
home environment.