Biohazard exposure. Obey safe handling protocols during contact with sample bottles and sampler
components.
Multiple hazards. The technician must make sure that the equipment operates safely and correctly
after maintenance procedures.
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or repaired,
contact the manufacturer.
7.1 Clean the instrument
Fire hazard. Do not use flammable agents to clean the instrument.
Do not clean the controller compartment heater with liquids of any kind.
If water is not sufficient to clean the controller and the pump, disconnect the controller and move the
controller away from the sampler. Allow sufficient time for the controller and pump to dry before the
parts are re-installed and put back into service.
Clean the sampler as follows:
• Refrigerator—clean the condenser fins and coils as needed with a brush or vacuum.
Note: The controller sets the temperature of the evaporator for frost-free operation. Do not use a mechanical
device or other procedure to increase the speed of a defrost cycle.
• Sampler cabinet and tray—clean the internal and external surfaces of the sampler cabinet with a
damp cloth and mild detergent. Do not use abrasive cleaners or solvents.
7.2 Replace the desiccant
A desiccant cartridge in the controller absorbs moisture and prevents corrosion. Monitor the
desiccant color through the window. Refer to
green, replace the desiccant.
1. Unscrew and remove the desiccant cartridge. Refer to
2. Remove the plug and discard the spent desiccant.
3. Fill the desiccant tube with fresh desiccant.
4. Install the plug.
5. Apply silicone grease to the O-ring.
6. Install the desiccant tube in the controller.
20 English
W A R N I N G
W A R N I N G
N O T I C E
C A U T I O N
N O T I C E
Figure
13. Fresh desiccant is orange. When the color is
Figure
13.