SAfETY PRECAUTIOnS bY USIng ThE CREAM :
• In case of allergy, stop to apply the cream.
• Do not swallow.
• Do not leave with the range children.
• In case of contact with eyes rinse with water.
PRESEnTATIOn Of ThE bELT :
• Front side (photo III. A, page 4) :
1 - Central fasteners 2 - Wire storage 3 - Connection terminals 4 - Scratch strip
• Electrodes side (photo III. B, page 4) :
4 built-in electrodes
STEP 1 : COnnECTIng AnD fITTIng ThE bELT.
• Check your MULTISPORTPRO unit is STOPPED
• Take the (CEV8) connector (photo I. B, page 3) and connect it :
- To the belt connectors per groups of colour (photo III. C page 4).
- Then to the unit ( photo III. D, page 4).
• Apply a thick layer of electro conductive massage cream onto the electrodes built into the belt (domed part) (photo III E page 4)
• Place the belt or 2 half-belts onto the zone(s) to be stimulated, and adjust them using "scratch" strips to ensure the built-in elec-
trodes are properly in contact with the skin
Photos IV page 4 : A - stomach, B - back, C - arms, D - thighs, E - calf, F - back.
• Move the belt slightly on left or right (with device powered off ) to feel the contractions on the abdominal area.
If an electrode hurts or if you feel any tingling, stop the device, and check the electrode is properly smeared with cream.
TIP :
• Fitting the full belt :
To well-positioned the full belt at best, cross the 2 strips to flip the 2 ends on the outer scratch part (photos V. A page 4).
• Fitting the half-belt on the arm :
Adjust at the belt mid-size, according to the outline of your arm before positioning it, then start over if necessary to ensure
proper adjustment (photos V. B page 4).
STEP 2 : STARTIng ThE bELT fUnCTIOn
Once your unit is connected, install yourself comfortably and start your unit using the procedure "Starting the unit" detailed
previously in paragraph II.
STEP 3 : STOPPIng ThE bELT fUnCTIOn
Once the program is completed, disconnect the unit from the (CEV8) connector, and the latter from the belt.
Remove the excess cream on the built-in electrodes using a wipe dampened with water, then dry them.
18
IV. REcOMMEnDaTIOnS OF USE
The unit is recommended for physiotherapists for functional re-education.
• You can restart a program's cycle on another muscle zone or change the program while remaining on the same muscle zone,
without exceeding however 40 min use per muscle zone and per day.
• For optimum use, refer to the diagrams shown in this leaflet.
• Clean the unit and its cables with a cloth dampened with a disinfecting solution (Mercryl type) diluted in water at 1/10
• The adhesive electrodes can be rehydrated before and after use with a disinfecting agent (e.g. Hexamedine).
• The adhesive electrodes must be repositioned on their support, and stored in a fresh storage between +5°C and +10°C.
• The adhesive electrodes have a lifetime of about 40 uses, you can order them directly from the online shop www.sport-elec.com,
or from your reseller.
PRECAUTIOnS Of USE :
• Do not use the unit on your face
• Do not leave within reach of children
Ask for advice from your physician :
- if you have electronic or metallic implants (pins, etc.)
- in case of skin condition (wounds,...)
wARnIngS :
• Do not use the device by driving or by working on a
machine.
• Do not use this electro stimulation unit in a damp environ-
ment, a bathroom, a sauna, or close to electric machinery
connected to mains and earthed, or even piping earthed.
• The simultaneous connection of a patient to a high
frequency surgical unit may cause burns at the points of
contact of the stimulator's electrodes, and the stimulator
may be damaged.
• Operation in close proximity of a short wave therapy unit
may cause instability in the stimulator's output power.
RECOMMEnDATIOnS :
Please apply regulations enforced in your
country when disposing your product at its
end of life.
Therefore, please discard it in a location planned to this effect,
ensuring it is disposed in an environment-friendly manner.
cUSTOMER SUPPORT: +33 ( 0 ) 2 32 96 50 50
th
.
COnTRA-InDICATIOnS :
Do not use SPORT-ELEC® :
• on the heart area
• If you carry a pacemaker
• If you are pregnant
• during first 6 weeks after baby birth
• In case of neuromuscular diseases
• In case of hemorrhagic diseases
RECOMMEnDATIOnS fOR MAInTAInIng AnD
CLEAnIng ThE bELT :
IMPORTANT : NEVER wash in machine and under running
water. The dryer and ironing facilities are forbidden.
• Do not dry-clean
• Do not expose to bleach
• Clean the belt using a cloth dampened with a detergent,
then wipe with a dry cloth.
• Clean the unit with a tissue dampened with a disinfecting
solution (Mercryl type) diluted with waterat 1/10th.
• Clean the electrodes with a mix of water/hypoallergenic
detergent, then rinse thoroughly
RADIATED ELECTRIC fIELD :
EN55011 / EN60601- 2- 10
Electric field at 3m.
Frequency (MHz)
(dB(μV/m)
200,0
< 30
400,0
< 30
600,0
< 30
800,0
< 30
1000,0
< 30