6. Error messages/ troubleshooting
If an error occurs during a measurement, the measurement is discon-
tinued and a corresponding error code is displayed.
Error No.
ERR 1
ERR 2
ERR 3
ERR 5
HI
LO
*If this or any other problem occurs repeatedly, please consult your doctor.
Possible cause(s)
The systolic pressure was determined
but afterwards the cuff pressure fell
below 20 mmHg.
This situation can occur, for example,
if the tube becomes unplugged after
the systolic blood pressure has been
measured. Additional possible cause:
No pulse has been detected.
Unnatural pressure impulses influence
the measurement result. Reason: The
arm was moved during the measure-
ment (Artifact).
Inflation of the cuff takes too long.
The cuff is not correctly seated or the
hose connection is not sealed.
The difference between systolic and
diastolic is excessive. Measure again
carefully following proper cuff proce-
dures and ensure measurement under
quiet conditions. Contact your doctor
if this message occurs frequently.
The cuff pressure is too high. Relax
for 5 minutes and repeat the
measurement.*
The pulse is too low (less than 40).
Repeat the measurement.*