multipiezo
11.1.3 Guide and the Manufacturer's Declaration - Electromagnetic
Immunity
11.1.3.1 Power Connection BC Input
multipiezo is designed to operate in the electromagnetic environment specified below.
The purchaser or user of multipiezo should ensure that it is used in such an environment.
Essential EMC
Phenomenon
standard or test
Electrical fast
transient/burst
IEC 61000-4-4
a) l) o)
Pulses
Differential
IEC 61000-4-5
mode
a) b) j) o)
Pulses
common
IEC 61000-4-5
mode
a) b) j) k) o)
Conductive
disturbances
IEC 61000-4-6
induced by RF
fields
c) d) o)
Voltage dips
IEC 61000-4-11
f) p) r)
Voltage
interruptions
IEC 61000-4-11
f) i) o) r)
62
Immunity test values
method
±2 kv on contact
100 KHz repetition
frequency
± 0.5 kv, ± 1 kv
± 0.5 kv, ± 1 kv, ± 2kv
3 V
m)
0.15 MHz - 80 MHz
6 V
in the ISM bands
m)
between 0.15 MHz and
80 MHz
80 % AM at 1 KHz
0% UT; 0,5 cycle g)
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % UT; 1 cycle e
70 % UT; 25/30 cycle
Single phase: at 0°
0% UT; 250/300 cycle
Environment Guidance
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
Portable and mobile RF
communication devices
should not be used near any
part of the product, including
cables, except when they
respect the recommended
n)
and calculated distances from
e)
the equation applicable at the
frequency of the transmitter.
The quality of the network
voltage should be that of a
typical commercial or hospital
environment.
h)
The quality of the network
voltage should be that of a
h)
typical commercial or hospital
environment.
Electromagnetic