LIGHT
(Ref. 3 inside cover)
Depending on the type of treatment
that needs to be carried out, 3 possible
options can be selected from the "light"
list:
• If the AUTO option is selected, the
LED light on the handpiece nose
cone lights up when the pedal is
pressed and automatically switches
off 3 seconds after the pedal is
released;
• If the ON option is selected, the LED
light on the handpiece nose cone
stays lit, regardless of the pedal.
The light turns off 100 seconds after
the last time the pedal is pressed,
and the option switches from ON
to AUTO;
• If the OFF option is selected, the
LED light on the handpiece nose
cone stays off.
NOTE: Adopting the "MANIPOLO
ABLATORE LED" ("LED SCALER
HANDPIECE" - cod. 03120142) equipped
with the "Front cone without light" (cod.
03020171), this function will not be
available.
The position of the LED light on the
nose cone of the "MANIPOLO ABLATORE
LED" ("LED SCALER HANDPIECE" - cod.
03120142) can be adjusted as follows:
1.
Hold the handpiece body and
slightly unscrew the metal ring
at the base of the nose cone by
turning it counter-clockwise;
2.
Rotate the nose cone so that the
LED light is in the desired and
necessary position;
3.
To fix the position, screw the metal
ring in a clockwise direction.
USE
perio/scaler
perio/scaler
restorative
restorative
19