TROUBLESHOOTING
Symptom
The display does not light up when
the unit is switched on
Handpiece without pressure or with
a low pressure when the foot switch
is activated
Pressure not adjustable
Pressure reduction during operation
Handpiece without pulse when the
foot switch is actived
Pressure too high
64
Solutions
Check the different electrical connections
Check the mains
Check/replace the fuses and replace them only with the type indicated on the back of the unit
Send the unit to your authorized EMS repair center
Check the different air connections
Check the pressure available for the unit
Check that the manometer is indicating more than 1.5 bar
Check the compressor
Check the connection between the handpiece and unit
Test the unit with another handpiece
Send the unit to your authorized EMS repair center
Check the different air connections
Check the pressure available for the unit
Check that the manometer is indicating more than 1.5 bar
Check the compressor
Send the unit to your authorized EMS repair center
Check the different air connections
Check the compressed air supply
Contact your authorized EMS repair center
Check that the "Validate" indicator light is on
Check the connection of the foot pedal
Check the different connections of the unit
Check the different supplies of the unit
Test the unit with another handpiece
Send the unit to your authorized EMS repair center