Register INPUT REGISTERS
function READ_INPUT_REGISTERS (0x30001)
Address
Type
0
int
3
Boot
4
bit
6.21
Temp. protection
AB69601en
Connection parameters:
IP address: the same as that of the device
Port: 502
Multiplier
10
-
-
The device is factory-fitted with a sample protection system providing
temperature protection. If any of the elements responsible for main-
taining the set temperature are damaged, or the user inadvertently
initiates a temperature setting, the set protection function will acti-
vate.
Temperature protection class
By default, the incubator units are equipped with protection class 2.0
and 3.1 in accordance with DIN 12880. The user can then program
the temperature value of the upper protection limit. If the set temper-
ature limit is exceeded, the heating system shuts down.
For class 3.1: Once the temperature returns to within the allowed
range, the device resumes operation.
The figure below shows how this works.
02.2023 / Rev. --
Description
Temperature from the main sensor
Open door
b0 – door alarm
b1 – upper temperature alarm
b2 – lower temperature alarm
b3 – over-protection
b5 – main sensor error
b7 – protection sensor error
b8 – temperature sensors error
b10 – hardware error
b11 – MRW error
Operation
61 / 82