Plug the wall adapter into the wall outlet.
To use a USB port for charging:
Place the M Waterproof Battery on a flat surface.
Connect the micro USB end of the power supply cable into the M
Waterproof Battery.
Insert the USB jack into a USB port (on a computer, in a car, etc.).
Please Note: After use, wipe the M Waterproof Battery with a dry cloth to
remove moisture or debris before charging.
Please Note: You may charge the M Waterproof Battery while the
processor is connected, but the sound processor will be powered
off during charging. Do not wear the M Waterproof Battery while
charging.
Understanding the M Waterproof Battery LED (light-emitting
diode) Indicators
The M Waterproof Battery has different indicator lights that are located
within the M Waterproof Battery.
The LED has 3 color options:
•
Blue: M Waterproof Battery is charging�
•
Green: M Waterproof Battery is fully charged�
•
Red: Fault condition�*
*Unplug the M Waterproof Battery and then try to charge again. If the
red LED illuminates again, contact your cochlear implant professional.
Understanding the M Waterproof Battery Audible Alarm
The M Waterproof Battery produces a constant, audible alarm when
the headpiece is unlocked from the implant in order to warn caregivers
when the headpiece becomes unlocked. The audible alarm is a
programmable feature that can be enabled or disabled by your cochlear
implant professional.
Streaming Audio while using the M Waterproof Battery
When the processor is connected to a Bluetooth-enabled device,
Phonak Roger transmitter, Phonak PartnerMic, or Phonak TV Connector,
you will be able to stream audio from that device directly to your sound
processor when it is in the M Waterproof Battery.
46