• The codes below are printed on batteries
containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.
12. Technical specifications
Model
Medel SENSE
Type
GCE602
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the upper arm
Measurement
Cuff pressure 0-300 mmHg,
range
systolic 50-280 mmHg,
diastolic 30-200 mmHg,
pulse 40-180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
diastolic ± 3 mmHg,
pulse ± 5% of the value shown
Measurement
Max. permissible standard deviation
uncertainty
according to clinical testing:
systolic 8 mmHg /
diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
L 134 mm x W 103 mm x H 60 mm
Weight
Approximately 367 g (without batteries,
with cuff)
Cuff size
22 to 42 cm
Noise
≤54dBA
Current
≤450mA
Service life
20.000 measurements
Permissible
+10 °C to +40 °C, 10 to 85% relative hu-
operating
midity (non-condensing), 800-1050 hPa
conditions
ambient pressure
Permissible
-20 °C to +55 °C, 10 to 90% relative hu-
storage conditions
midity, 800-1050 hPa ambient pressure
Power supply
4 x 1.5 V
Battery life
For approx. 300 measurements, de-
pending on levels of blood pressure
and pump pressure
Classification
Internal supply, IP21, no AP or APG,
continuous operation, application part
type BF
The batch number is located on the device or in the battery
compartment.
Technical information is subject to change without notifica-
tion to allow for updates.
61
AA batteries