Descargar Imprimir esta página

smartheart 01-742T Manual De Instrucciones página 32

Publicidad

Idiomas disponibles

Idiomas disponibles

ELECTROMAGNETIC COMPATIBILITY
Phenomenon
Electrical fast transients/
burst
Surges
Line-to-line
Surges
Line-to-ground
Conducted disturbances
induced by RF fields
Voltage dips
Voltage interruptions
32 • ENGLISH
Table 4 — INPUT A.C. POWER PORT
Basic EMC Standard
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Immunity test levels
Home healthcare environment
±2 kV
100kHz repetition frequency
±0.5 kV, ±1 kV
±0.5 kV, ±1 kV, ±2 kV
3V, 0.15MHz-80MHz
6V in ISM and amateur radio bands between 0.15MHz and 80MHz
80%AM at 1kHz
0% UT; 0.5 cycle
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0% UT; 1 cycle and 70% UT; 25/30 cycles, Single phase: at 0º
0% UT; 250/300 cycles

Publicidad