ERROR CODES
Code
Batteries are missing, inserted incorrectly
Monitor cannot
or flat.
be switched on
Cuff connector plug is incorrectly positioned
Cuff will not
in the connector socket of the monitor.
inflate
Pulse could not be detected correctly.
E1
E2
The pulse could not be detected.
E3
The cuff was applied too firmly or too
loosely.
Errors occur during the measurement;
E4
excessive movement
Inflation pressure is higher
E5
than 300 mmHg.
E6
There is a system error.
The batteries are almost empty.
Implausible measured values often occur
Implausible
due to inappropriate handling of the device
measure values
or mistakes during the measuring procedure.
Switch the device off if an error symbol appears. Check for all possible causes and note the instructions for self-measurement.
Relax for a minute and then take the measurement again.
In such cases, repeat the measurement. Ensure that you do not move or speak. If necessary, reinsert or replace the batteries.
Possible Causes
Corrective Action
Check batteries and insert four identical new batteries
if necessary.
Check the connection between the cuff connector plug
and socket.
Check whether the cuff has been applied correctly. Do
not talk or move during the measuring procedure.
Check whether the cuff has been positioned correctly.
Do not talk or move during the measuring procedure.
Apply the cuff so that two fingers fit between the cuff
and the upper arm.
Air tube is not correctly connected to the monitor.
Check that the connector plug is correctly positioned. If
this error occurs frequently, you should use a new cuff.
Please repeat the measurement. Make sure that the air
tube is properly inserted.
Please rest for at least 1 minute and take the
measurement again.
Contact customer service if this error message appears.
Replace batteries.
Please repeat the measurement according to the
instructions.
ENGLISH • 29