ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacture's declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of this monitor should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz) magnetic Field
IEC 61000-4-8
32 • ENGLISH
Table 2 — IMMUNITY
For all ME EQUIPMENT and ME SYSTEMS
IEC 60601 test level
±8 kV contact
±15 kV air
30 A/m
50Hz/60Hz
Compliance levels
±8 kV contact
Floors should be wood, concrete or ceramic tile.
±15 kV air
If floor are covered with synthetic material, the
relative humidity should be at least 30%.
Power frequency magnetic fields should be at
30 A/m
50Hz/60Hz
levels characteristic of a typical location in a typical
commercial or hospital environment.
Electromagnetic environment
guidance