Error Messages - Hartmann Veroval Instrucciones De Uso

Tensiómetro para el brazo
Ocultar thumbs Ver también para Veroval:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

English
GB
If the data transfer is unsuccessful, an error message is displayed on the blood
pressure monitor, as shown in Fig. 2.
In that case, remove the connection to the PC and start the data transfer again.
The blood pressure monitor switches off automatically after 10 seconds of inac-
tivity and when communication with the PC is interrupted.

8. Error messages

Specific
Error
Monitor cannot
be switched on
Cuff will not
inflate
E  I
E2
Fig. 2
Possible causes
Batteries are missing,
inserted incorrectly or
flat.
Power supply unit not
correctly connected or
faulty.
Cuff connector plug is
incorrectly positioned in
the connector socket of
the monitor.
Wrong cuff type con-
nected.
Pulse could not be
detected correctly.
The pulse could not be
detected.
Remedy
Check batteries and insert four
identical new batteries if nec-
essary.
Plug the power supply unit into
the connector socket on the
back of the monitor.
Check the connection between
the cuff connector plug and
socket.
Verify that only an approved
Veroval
®
being used.
Check whether the cuff has
been applied correctly. Do not
talk or move during the measur-
ing procedure.
Check whether the cuff has
been positioned correctly. Do
not talk or move during the
measuring procedure.
88
cuff and connector are

Publicidad

Tabla de contenido
loading

Tabla de contenido