Descargar Imprimir esta página

InterMed pinkwave VISTA APEX Manual De Instrucciones página 13

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

• Use only the charger that is provided with the device, cable type for this device
is NEMA1-15 to IEC 320 C7 (non-polar). The use of any other charger can result
in damage to the battery pack.
• Condensation resulting from the device being transferred from a cold to a warm
environment may be a potential risk. Never begin operating the device until it
has reached the ambient temperature.
• There are no user-serviceable items in this handpiece or charger.
No modification of this equipment is allowed.
• In order to avoid electric shock, do not introduce any objects into the
device or remove the device enclosure.
• Should you have any reason to suspect the safety of the device to be compromised,
the device must be taken out of operation and labeled accordingly to prevent
third parties from inadvertently using a possibly defective device. Safety may be
compromised, e.g., if the device malfunctions or is noticeably damaged.
• Keep solvants, flammable liquids, and sources of intense heat away from the
device as they may damage the plastic housing of the device, the seals or the
operating buttons.
• According to IEC 60601-1, this device must not be used in the presence of
flammable mixtures.
• Do not allow any cleaning agents to enter the device during cleaning as they
could cause an electrical short or a dangerous malfunction.
• Only Vista Apex may open the device housing and repair the device.
• UV emitted from this product. Eye or skin irritation may result from exposure.
Use appropriate shielding.
• IR emitted from this product. Do not stare at operating lamp.
• To avoid the risk of electric shock, this equipment must only be connected to
a supply mains with protective earth.
• Use only components and accessories listed in the instructions associated with
the device. Failure to do so will void the warranty, may decrease the performance,
and may lead to unsafe operation. Other cables may negatively affect EMC
performance of the device.
• Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally.
13

Publicidad

loading

Productos relacionados para InterMed pinkwave VISTA APEX

Este manual también es adecuado para:

90734