• PinkWave™ must not be used in patients, or by users, with heart pacemaker
implants who have been advised to be cautious with regard to their exposure to
small electrical devices.
• This device has been tested and found to comply with relevant EMC regulations
and standards. These limits are designed to provide reasonable protection against
harmful interference in a typical medical installation. The device generates
radio frequency energy and, if not installed and used in accordance with these
instructions, may cause harmful interference to other devices in the vicinity.
However, there is no guarantee that interference will not occur in a particular
installation. If this device does cause harmful interference with other devices,
which can be determined by turning the device off and on, the user is encouraged
to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving device.
Avoid stacking with other equipment.
• Increase the separation between the devices.
• Connect the device into an outlet on a circuit different
from that to which the other device(s) are connected.
• Consult the manufacturer for help.
• Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the PinkWave™, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
OPERATING CONDITIONS AND
SAFETY CONSIDERATION
Heat Generation
The metal parts of the PinkWave™ Curing Light do not overheat to the point of discomfort
or injury during standard operating durations (i.e. five consecutive standard mode, ramp
mode or boost mode cures). However, care should be taken to allow the product to
completely cool to room temperature between patients (approximately two minutes),
to ensure overheating does not occur.
CAUTION: The PinkWave™ consists of high powered LEDs in the range
of 375 – 900nm. The device's intended use is within the oral cavity and
is to be used with the protective light shield.
14