Infection Control Measures
CAUTION: To prevent cross-contamination, a disposable plastic sleeve must
be used over the PinkWave™ with each use. A low-density polyethylene plastic
disposable barrier covers the entire unit and provides a hermetically sealed
barrier between the handpiece and patients. The disposable barrier limits patient-
to-patient contamination. Discard used barrier sleeves after each patient.
Routine Maintenance
Periodically check the lens for cured dental resins. If necessary, carefully remove any
adhered resin using a non-diamond dental hand instrument. The barrier sleeve helps keep
dental materials from adhering to the surface of the lens.
TROUBLESHOOTING GUIDE
If the suggested solutions do not rectify the problem, please call Vista Apex
at 877-273-9123 (Toll Free).
PROBLEM
POSSIBLE SOLUTION
Light will
1. Check the unit's indication light. If red, place the handpiece on
not turn on
2. Remove the battery pack and inspect for any defects if none are found
Light is
1. Verify the unit provides adequate optical power using a radiometer.
not curing
2. Verify that the appropriate curing mode is selected.
properly
3. With the handpiece turned off, and the battery pack removed,
4. Check expiration date of composite/resin.
Batteries
1. Remove the battery pack and inspect for any defects, if none are found
will not
charge
2. Make sure the battery pack is properly connected to the handpiece.
3. Make sure the charger is plugged in and verify the outlet is receiving power.
4. Lights on charger should pulsate.
the charging base.
reconnect the battery pack. A beep should be heard when the battery pack
is inserted.
inspect the lens for residual cured composites/resins.
reconnect the battery pack.
9