Descargar Imprimir esta página

InterMed pinkwave VISTA APEX Manual De Instrucciones página 8

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Transillumination Mode
The transillumination mode can be executed by toggling through modes until all lights are
illuminated and pressing the ON/OFF button. The transillumination light is turned off the
same way, by pressing the ON/OFF button.
Disinfection / Sterilization
The PinkWave™ Curing Light is provided non-sterile. There are not any special accessories
needed to sterilize the PinkWave™ Curing Light parts.
Clean the PinkWave™ Curing Light and Light Shields with an approved surface disinfectant
solution such as a quaternary ammonium compound product that contains 20% alcohol or less.
Wipe, do not spray, solution and follow the manufacturer's cleaning recommendations.
Prevent liquids from entering openings on the unit.
For Protective Eyewear: Clean Protective Eyewear lens with water only. Use a soft cotton
or microfiber cloth and gently pat lens dry. Do not use abrasive cleaning compounds,
solvents, ammonia or alkaline cleaners.
For PinkWave™ Paddle: DO NOT autoclave the paddle. Autoclaving the paddle may cause
damage. For cleaning the paddle in between patients, apply an appropriate medical grade
surface disinfectant (e.g. Cavicide) to all surfaces of the paddle using a lint-free wipe per the
disinfectant's instructions.
NOTE: The Light Shields can also be autoclaved. DO NOT autoclave the
handpiece, battery pack, charging base or PinkWave™ Paddle.
CAUTION: DO NOT use abrasive material such as scouring powder,
organic solvents, or solvent-based cleaning fluids. In case of severe
contamination, gently clean the device by using diluted alcohol.
Store the device in the box if it is not to be used for an extended period of time.
Disassembly is not required when cleaning the handpiece and charging base.
8
CAUTION: DO NOT immerse the unit or unit parts in
solutions. Use of solutions other than those recommended
may damage plastic parts and will void product warranty.

Publicidad

loading

Productos relacionados para InterMed pinkwave VISTA APEX

Este manual también es adecuado para:

90734