7 Reprocessing
34
2.
If they are still dirty, repeat the process.
Ä
For further reprocessing, the instrument should be dry and
free of residue.
3.
Blow the instrument out with max. 3 bar.
4.
Pack the instrument in packaging material suitable for
sterilization and storage. e.g. paper/plastic composite
packaging.
5.
Conduct sterilization  [→ 37].
7.3
Manual cleaning and disinfection
IMPORTANT: Manual conditioning is possible in exceptional cases
if the national/regional regulations are followed. The national/
regional regulations are to be checked before.
NOTE!  Condition immediately, or at the latest, one hour after
treatment.
NOTE!  Never clean in an ultrasound bath!
Operating Instructions BACS125 / BACS126