Descargar Imprimir esta página

B.A. International BACS125 Manual Del Operador página 40

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 142
7 Reprocessing
38
Operating Instructions BACS125 / BACS126
Overpressure:
2.04 bar (29.59 psi)
Temperature:
134 °C (274 °F)
Holding time:
3 min.
Steam sterilizers that comply with the requirements of either
EN 13060, class B or EN 13060, class S and are also suitable for
the sterilization of ultrasonic handpieces and ultrasonic tips are
approved.
NOTE!  Do not exceed 140 °C (284 °F), even during the drying
phase.
After sterilizing
1.
Remove the instruments from the steam sterilizer immediately.
CAUTION!   The instruments are hot. Risk of burns!
NOTE!  Do not attempt to accelerate the cooling process by
immersing the instruments in cold water. This can damage
your instruments!
63 83 066  D3275.201.08.03.02

Publicidad

loading

Este manual también es adecuado para:

Bacs126