CAUTION
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
• Remove all parts from the disinfector as soon as the disin-
fection process is finished and allow them to dry.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.
DRYING:
• After the disinfection process is complete, place all parts on
a dry, clean and absorbent surface and allow them to dry
completely. Or leave all individual parts in the closed
thermal disinfector for max. 24 hours until the next use.
4.6 Care of the connection tube
Dry the connection tube after every inhalation session:
• Connect the connection tube to the compressor.
• Switch on your compressor.
• Let the compressor continue to run until all the moisture in
the tube has been removed.
4.7 Inspecting
Inspect all product components after each cleaning and disin-
fection. Replace any broken, deformed or seriously discoloured
parts.
4.8 Drying
After each cleaning and disinfection, place all product compon-
ents on a dry, clean and absorbent surface and let them dry
completely.
4.9 Storage
Store this product as described below:
• Wrap all individual components in a clean, lint-free cloth
(e.g. a tea towel).
– 26 –