Cleaning with disinfection
Cleaning and disinfection can be carried out in a single cycle
using a chemical preparation process.
To ensure safety when handling chemicals, follow the instruc-
tions for use of the disinfectant.
Mechanical
cleaning with
disinfection:
EQUIPMENT:
The method has been validated in Europe using:
– Alkaline cleaning agent:
Dr. Weigert neodisher
centration: 0.5%)
– Deionised water
– Cleaning and disinfection device:
Steelco DS800 in conformance with
DIN EN ISO 15883-1 and 15883-2
Info: If a different alkaline cleaning agent is
used, it may also be necessary to use a
neutralising agent. Follow the recommendations
of the manufacturer of the chemical.
PROCEDURE:
Programme for cleaning and disinfecting
according to manufacturer's instructions.
DRYING:
Ensure that there is no residual moisture
remaining in the components.
• Shake the water out of all of the parts.
• Place all parts on a dry, clean and absorbent
surface and allow them to dry completely.