Preparation For Use/Operating Instructions; Checks Before Use; Connect Tubing - Medela Thopaz+ Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 303

7 Preparation for use/operating instructions

WARNINGS
– For use only by medically trained and qualified persons who have been adequately
trained in the use of Thopaz + . Wear gloves for all operations. "Pressure" generally
implies "negative pressure".
– Adapt pressure settings for pediatric patients.
CAUTIONS
– Do not use sterile accessories when the sterile packaging is damaged.
– Non sterile and reusable accessories must be cleaned and disinfected according to
chapter 13 – General cleaning guidelines.
– Do not switch Thopaz + on before first charging the battery.
7.1

Checks before use

– Check the Thopaz + system before use for damage of the power cord or plug, obvious
device damage or safety defects and proper functioning of the device.
– Check that the seal is correctly placed in the vacuum port of Thopaz + . If you do not
know where the vacuum port is placed please refer to chapter 4 – Overview.
– Make sure that the rechargeable battery is recharged in case Thopaz + is operated in
battery mode.
– Check all accessories prior to use:
– Canister for cracks, brittle and flawed spots.
– Tubing for cracks, brittle areas and that connectors are firmly attached.
Replace if necessary.
7.2

Connect tubing

1.
2.
1.1 Open external packaging.
1.2 Keep patient connector in
internal bag for hygienic
reasons.
2.1 Insert the smaller of
the two connectors
(connection to pump)
horizontally in the
direction of the arrow.
Preparation for use/operating instructions
| 25

Publicidad

Tabla de contenido
loading

Tabla de contenido