Carry Out Functional Check - Medela Thopaz+ Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 303
7.5
Carry out functional
check
Safety related tip
The acoustic alarm "Leak in system – 301" is deactivated for approximately 6 minutes
after Thopaz + is switched on.
Safety related tip
Everytime a new patient is connected, it is recommended to carry out the functional
check.
28 |
Preparation for use/operating instructions
1.
2.
Thopaz + is in Standby and is ready for use with the factory
3.
settings. (To change factory settings please refer to chapter
10 – Changing settings)
Connect the Thopaz + system to the patient catheter (drain)
4.
according to hospital guidelines. To start the therapy press
"On".
1.1 Seal off the conical
connector of the patient
tubing with your thumb
(through the packaging).
The tubing must not be
kinked otherwise the
measuring tube will be
closed.
2.1 Press "OK"
Test passed:
Thopaz + is ready for use.
Test failed:
Leak in the system:
– Check tubing connection.
– Check seal is correctly
placed.
– Check pressure relief valve
for tight connection.
– Replace canister and/or
tubing if necessary.

Publicidad

Tabla de contenido
loading

Tabla de contenido