5. Attach the sensor to the positioner. The sensor will only fit into the
holder one way.
6. Position the sensor appropriately in the part of the mouth to be X-rayed
(refer to specific instructions for use).
Important:
The sensor must be positioned with the sensitive area facing
the source of the radiation. The sensitive side of the sensor is
marked with the DEXIS logo.
NOTE: Use of a sensor positioning device is recommended to ensure that the
sensor is positioned at right angles to the radiation beam and parallel
to the tooth. Positioning devices are supplied with the DEXIS system.
Portability
DEXIS Sensor can be easily moved from one dental chair to another. Once
disconnected from the USB port of the computer, the sensor may be positioned near
a second dental chair and connected to a USB port present nearby.
The USB technology allows easy connection and disconnection of DEXIS Sensor
even if the computer is switched on, unless the PC presents particular limitations.
No activation or deactivation procedure is necessary with DEXIS Sensor, just plug
in and out.
Hygiene
CAUTION
After each patient, the sheath must be removed and the sensor must
be cleaned and disinfected by CaviWipes; h
disinfected in an autoclave or with a cleaning agent after each patient.
Store disposable sanitary sheaths in a clean, dry place not exposed to sunlight or
UV rays.
Make sure that used sheaths are disposed of as infected waste which is potentially
biologically hazardous.
Important:
Disinfect the image sensor before its first use and whenever
there is a risk of contamination.
Please follow the disinfecting and cleaning instructions in order to avoid damaging
the sensors. Disinfection of the X-ray image sensor at the end of each day is
recommended even when sheaths are used.