WARNING
Do not autoclave sensor. Autoclave sterilizers will permanently
damage the DEXIS Sensor.
Recommended Disinfecting Solutions
Decontaminate the sensor, cable and holders in accordance with CDC – or your
country's (e.g., OSAP) standards – for infection control. DEXIS recommends
wiping the DEXIS Sensor with one of the following cleaning agents.
CaviCide
CaviWipes
ProSpray
Sani-Cloth
DO NOT:
Clean the sensor using non appropriate tools.
Maintenance
DEXIS Sensor does not require any special maintenance other than regular
cleaning and disinfection. It is recommended use DEXIS Software Manager for
automatic software updates.
Go to dexis.com/downloads for more information on using DEXIS Software
Manager. DTX Go may also be used for managing your software updates.
Image Quality Assurance
Image quality of the DEXIS Sensor depends on several factors:
the quality of the X-ray source (kV, focal spot size, distance)
the alignment of the X-ray source to the anatomic region
the applied X-ray dose / exposure time
the settings of the computer monitor
It is recommended that you establish a procedure for periodic review of the image
quality. If image quality is not satisfactory, or degrading, please check the
contributing system parts as outlined below: