DEXIS
Environmental Conditions
USAGE
DEXIS Sensor is not suitable to
be operated in oxygen rich and/
or explosive environments
TRANSPORTATION AND
STORAGE
Transport in the supplied
protective package
EU Classification
Protection against water/matter – IP 68
Compatibility with Radiographic Generators
DEXIS Sensor is generally compatible with any dental X-ray unit and generator
capable of supplying the required range of exposure times and doses.
Follow the instructions of the X-ray generator to set the desired dose range.
The radiation of X-ray tubes is controlled by the settings of:
Some controls allow for the modification of all of the above, some have fixed settings
for current and voltage.
Distance of X-ray Source to Sensor
There is a correlation between the distance of the X-ray source to the sensor and
the dose received by the DEXIS Sensor. The radiation that reaches the sensor
decreases with the square of the distance. If you double this distance, you receive
only 1/4 of the radiation.
6-2
Humidity
30% to
95%
10% to
95%
Class IIa medical device according to MDD
Exposure time (msec) or pulses
Voltage (kV or kVp)
Current (mA)
Air
Ambient
Pressure
Temperature
5° to 30°C
700 to
1060
hPa
-40° to 70°C
93/42/ECC