Device preparation
6.3.1
Preparing the electronic motor
Sterilize the electronic motor that drives the Conform Cannula Handpiece (the motor is not sterile on deliv-
ery). If the motor has already been sterilized: when removing the motor from the sterile packaging, ensure
that the sterile packaging is not damaged and that the sterility indicator confirms sterility (if no sterility
indicator is provided, the sterile packaging must at least show the date on which the shelf life of the sterile
item is due to expire).
6.3.2
Preparing the bottle holder
Insert the bottle holder stand for the suspension of the tumescent solution container into the stand holder
at the control unit and hang the container on it. The maximum load on the hook is limited at 2 kg.
Heavier loads than allowed (2 kg) on the bottle holder stand may cause the unit to tip over.
6.3.3
Preparing the pedal
Connect the pedal plug with the pedal socket at the rear of the control unit.
6.3.4
Preparing the Conform Cannula Handpiece
Assemble the sterilized Conform Cannula Handpiece, electronic motor and all the accessories according to
the following instruction.
1
2
1.
Liposuction cannula (optional)
2.
Cannula adapter with suction channel (REF 75705)
3.
Connecting piece of suction tube (REF 28987)
4. Transmission capsule (REF 75700)
5.
Suction pipe integrated in handle
Remove locking clamp.
10
5
3
4
Remove blocking ring.
6
6
7
8
6. Handle (REF 5107)
7. Slot opening for locking clamp
8. Locking clamp for blocking ring (REF 28556)
9. Electronic motor (REF 2101nou)
10. Blocking ring (REF 28554)
Unscrew cannula adapter from ...
Operating instructions LipoSurg, REF 31648, V48/18
LipoSurg
9
10