Descargar Imprimir esta página

Nouvag Lipo Surg Instrucciones De Uso página 49

Ocultar thumbs Ver también para Lipo Surg:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

LipoSurg
10
Malfunction and troubleshooting
Malfunction
Device doesn't work
No stroke movement at the
Conform Cannula Handpiece
No delivery of infiltration
solution at the cannula
Pedal is not functioning
If a fault cannot be rectified, please contact your supplier or an authorized service center. The addresses are provided on
the last page of the operating instructions.
Operating instructions LipoSurg, REF 31648, V48/18
Cause
Control unit is
not switched on
Power connection not estab-
lished
Incorrect operating voltage
Faulty fuse
Processor error
"LIPO" function is not select-
ed
Motor to drive cannula not
activated
Motor to drive cannula not
connected
Transmission capsule not
connected properly to the
motor
"INFIL" is not selected
Peristaltic pump is not
switched on
Tube set not properly inserted Insert tube set correctly
Tube set is blocked
Infiltration container is not
ventilated
Tube set is dripping
Roller clamp is closed
Pedal not connected
Control unit is not switched
on
Faulty operation
Solution
Set the
power switch "I/O"
to "I"
Connect the control unit to
the mains power supply
Check the mains voltage
Replace the fuse
Switch main switch "I/O" to
position "O", wait 10 sec. and
switch back to position "I".
Select function "LIPO"
Activate motor to drive can-
nula by pressing pedal
Insert motor plug into the
socket at the devices front
Press transmission capsule
firmly to the motor until it
engages.
Select function "INFIL"
Activate peristaltic pump by
pressing the pedal
(watch flow direction)
Replace tube set
Open air bleed valve at
tube set
Replace tube set
Open roller clamp all the way
it goes
Insert the pedal plug into the
socket at the devices back
Use mains switch, in the back
of the device, to switch on
Read operation manual care-
fully
Refer to operating instruction
7.1
Switching the device on and
off
6.2
Connection to the power
supply
6.2
Connection to the power
supply
9.1
Replacing the control unit
fuse
7.1
Switching the device on
and off
7.2
Operation elements
7.6
Operation with VARIO pedal
5.0
Device overview
6.3.4 Preparing the Conform
Cannula Handpiece
6.3.4 Preparing the Conform
Cannula Handpiece
7.2
Operation elements
7.6
Operation with VARIO pedal
6.3.5 Preparing the tubing set
6.3.5 Preparing the tubing set
6.3.5 Preparing the tubing set
6.3.5 Preparing the tubing set
6.3.5 Preparing the tubing set
6.3.3 Preparing the pedal
5.0
Device overview
7.1
Switching device on and off
7.6
Operation with VARIO pedal
EN
21

Publicidad

loading