LipoSurg
7.3
Infiltration "INFIL"
A
B
A)
Selected function "INFIL"
By pressing the "INFIL" –key, the "INFIL" -function (Infiltration) is displayed up left.
By pressing the pedal the display shows "INFIL in operation" up right.
B)
Setting the maximal by pedal retrievable volumetric displacement of the peristaltic pump.
The output limit of the tumescent solution can be regulated in 10 % steps, beginning with 10 %.
C)
Real-time display of the pump delivery rate, which is retrieved by pedal
Bar graph shows the percentage of flow rate of the peristaltic pump, retrieved by pedal.
The pump does not begin to operate until the motor has been activated by pressing the pedal.
7.4
Liposuction "LIPO"
A)
Selected function "LIPO"
By pressing the "LIPO" –key, the "LIPO" -function (Liposuction) is displayed up right.
By pressing the pedal the display shows «LIPO in operation» up right
B)
Setting range of the number of cannula strokes
Display of the range of possible cannula strokes.
C)
Regulating the cannula strokes
By pressing the pedal the Conform Cannula Handpiece is running with the preset value.
7.5
"COMBI-MODE"
INFIL
Selected function „INFIL & LIPO"
By pressing both keys at the same time "INFIL" & "LIPO" the "COMBI MODE" (Infiltration & Liposuction)
is displayed.
Set values for both of the functions just as mentioned above.
Operating instructions LipoSurg, REF 31648, V48/18
Display in operation mode.
A
B
B
C
LIPO
F
G
H