7.6
Operation with VARIO pedal
The pedal can be used in two different modes for not having to press the pedal permanently during infiltra-
tion or liposuction.
In Normal mode (variable) it is required to press or move the
pedal permanently to vary the infiltration up to the preset max-
imum value, while at liposuction the stroke of the cannula
works always with the preset stroke rate.
In ON/OFF mode the pedal behaves like a switch. By shortly
pressing the pedal, infiltration and liposuction is carried out with
the preset values. It is not necessary to keep pressing the pedal.
If the pedal is again pressed shortly, the device stops immedi-
ately.
7.6.1
Pedal mode selection
The switching of the pedal mode starts with the device
switched off.
 Press the "INFIL"-key and switch the device „ON", simulta-
neously.
 For switching from „Normal"-mode to „ON/OFF"-mode,
press the „LIPO" key.
When switching the device „OFF", the selected mode will re-
main at the next time when you turn it on again.
7.6.2
Normal mode (variabel)
Pedal...
... not pressed
... gently pressed
... pressed all the
way
7.6.3
ON/OFF mode
Pedal...
... not pressed
... shortly pressed
... shortly pressed
again
For safety reasons, the unit can only be operated by pedal.
16
Cannula stroke:
Peristaltic pump:
No cannula stroke
Pump off
Cannula runs with
Pump runs with the speed
maximal speed
shown by the bar graph
Cannula runs with
Pump runs with the maximal
maximal set speed
speed
(Bar graph shows all seg-
ments full).
Cannula stroke:
Peristaltic pump:
No cannula stroke
Pump is off
Cannula works with
Pump runs with preset value
preset value
No cannula stroke
Pump is off
Liposuction with ON/OFF pedal mode
Infiltration with ON/OFF pedal mode
Display after entering the pedal mode selection
Display after switching, by pressing the „LIPO"-key
Operating instructions LipoSurg, REF 31648, V48/18