All manuals and user guides at all-guides.com
D. Engineering units
Technical unit for gas, temperature and pressure can be set.
Press:
Select the parameter for changing, pressing
Confirm the selection pressing the key
Set
Day , Month or Year
Confirm changes pressing the key
E. CO diluted to Oxygen reference
The CO undiluted (0% O ) setting is standard set. Use the procedure below
if a different reference value is required.
Select and confirm the "O2 reference" option.
Press:
Select the parameter for changing, pressing
Confirm the selection pressing the key
Set
the reference
Confirm changes pressing the key
1. Switch the Instrument ON
Be sure that the boiler room is well vented and secure from CO.
Before entring in the boiler room check the ambient CO concentration.
1. Set the right fuel for the boiler
The fuels can be changed.
Up to 10 fuels can be set using GasConfig PC software and the USB cable.
Select and confirm the "--- FUEL ---" option.
Press:
Select the Fuel from a list pressing
Confirm the selection pressing the key
Press
ESCAPE
To pause the analysis press the
pump. Press
ESCAPE
2. Pressure Draft procedure
A. Internal pressure sensor
Differential pressure and draft can be measured using the internal
pressure sensor (P1 - P2) inlet. Do not exceed 200mbar input.
REMOVE THE PROBE FROM THE STACK
#
Press
DP
Press
to input the ambient temperature
Press
F1
to select the Internal Sensor input
Press
FI
"ZERO" to reset the display indication.
Insert the probe inside the standard
boiler holes (see figure). If the boiler do not have
them, insert the probe as shown in the drawing.
Wait for stabilization of the measurement.
Press
to save the measured value.
Value will printed on the report.
Press
F2
to exit to the previous menu.
6
F1
value, pressing
2
F1
value, pressing
OPERATIONS
Wait the 60 seconds countdown
for the autozero procedure.
DO NOT INSERT THE PROBE IN THE STACK
To mute the beep
(F2) to start the flue gas analysis
(F2) to restart the flue gas analysis.
ATTENTION:
button. This will stop the internal
- applicable for natural draft boiler
ENGINEER UNITS
> CO
ppm
Other
ppm
Temp.
°C
Pres.
hPa
ESCAPE
F1=Internal sensor
F2=External sensor
TaE
23.00
INT
EXT
DRAUGHT
Measure
1.0 mbar
Press [ENTER]
To save measure
ZERO
ESCAPE
D
–
–
–
2 x D
5