2.8. Safety information for the individual operating
phases
2.8.1 Transport
f Transport the device only as described in the Operating Instructions (see
requirements regarding transport and setting up in the Operating Instructions,
"Documents of equal applicability and required files", Page 5).
f In order to avoid corrosion on the device and subsequent damage to the device,
transport and store device only in the permissible temperature range and ambient
conditions (see requirements regarding the ambient conditions in the Operating
Instructions, "Documents of equal applicability and required files", Page 5).
2.8.2 Initial operation
f Ensure that this device is only operated by authorized and trained specialists.
f If the device has been stored in a cold environment or at high humidity, maintain a
drying or temperature adjustment time of approx. 4 hours (without voltage) at room
temperature before initial operation.
f Before connecting the device, check that the supply voltage and frequency are
correct at the point of installation (see electrical data in the Operating Instructions,
"Documents of equal applicability and required files", Page 5).
f The requirements regarding the location site and ambient conditions must be
observed (see requirements regarding the location site and ambient conditions in the
Operating Instructions, "Documents of equal applicability and required files",
Page 5).
2.8.3 Operation
f Ensure that this device is only operated by authorized and trained specialists.
f Keep unauthorized persons, such as patients and children, as well as animals away
from the unit.
f In all cases of doubt regarding the safety of the device, switch off and disconnect the
device and take suitable measures to prevent further use.
f Prior to connecting power or operation, check the device, the accessories and
protective equipment for any damage.
f Do not use damaged, non-functioning equipment or accessories; instead notify your
authorized service partner.
f In order to ensure the product safety and warranty services, the device must be
exclusively operated with the original accessories from Ivoclar, particularly the original
power cord. The user bears the risk when using non-approved accessories.
f Ivoclar recommends using only approved materials to ensure process reliability.
f Never bypass the safety equipment of the device or set it out of operation ("Working
areas, potential hazards and safety measures on the device", Page 8).
f Only operate the device when the build chamber is closed.
f To prevent damage to the device and a reduction of the device performance, observe
the cleaning requirements and cycles (see "Cleaning by the user", Page 12 in these
Short Instructions as well as the overview and instructions for the cleaning and
maintenance work in the Operating Instructions, "Documents of equal applicability
and required files", Page 5).
f Only operate the device unsupervised if the operating conditions for unsupervised
operation described below are fulfilled.
Unsupervised operation:
The device may be operated unsupervised, provided the national and local laws and
provisions allow for such action and provided that they are observed. Furthermore, the
requirements of the respective insurance company must be met.
f Never use the device if the build chamber is heavily soiled.
f Protect the device against unauthorized access.
11