5 Limitations on Use
Intended Life
A local risk assessment should be carried out based upon activity and usage.
Environment
Avoid exposing the device to corrosive elements such as acids, industrial detergents, bleach or
chlorine. The use of creams or lotions with this device should be used with caution as these can
cause the device to soften and overstretch or distort.
Keep away from sharp objects (such as jewellery,
fingernails).
Exclusively for use between -15 ˚C and 50 ˚C (5 ˚F to
122 ˚F).
6 Donning the Device
Before you begin
Fit the valve. (See Fitting the Valve section.)
Care must be taken during donning/
doffing not to damage the device
with fingernails, sharp jewelry or
the locking pin.
Do NOT pull or stretch the device.
1.
Invert the device so that its silicone side points
externally.
2.
Align the locking pin with the long axis of the
residual limb.
3.
Roll the device on the residual limb while releasing
any trapped air.
Note: The unperforated fabric should be below the
patella tendon.
After donning the device, allow the user to wear it for
10 minutes. If the user feels numbness, tingling or any
unusual sensation within 10 minutes, do the following:
4.
Doff the device, and wait until normal sensation
returns.
5.
Don the device again.
6.
If the user feels numbness, tingling or any unusual
sensation again, stop using the device.
7
938399PK1/3-0123