Power unit
For an optimum lifetime the power unit must
be connected to the mains voltages as often as
possible (e.g. every night). Charge the batteries
12 hours before use.
The batteries must be charged at least every
3rd month - otherwise there might be damages
due to self-discharge. It is recommended to test
the battery function at least once every year.
Recharging the battery:
• Remove the plug from the power system and
plug in the charging lead
• Plug the charger into the electricity supply
and switch ON. A yellow light on the charger
will light up
• Charge the batteries until the lamp on the
charger turns green
Technical Data
Battery - BA18
• Capacity 1.3 Ah - 24V DC
• Charging current is max. 500 mA
• IPX6
The batteries have a maximum life of 4 years
but this can vary depending upon the conditions
of use and charging practices
Battery charger - CH01
• Mains: 100 - 240VAC/50-60 Hz switch mode
• Charging voltage: 27.6 VDC
• Charging current: Max. 500 mA
• FRIWO Gerätebau GmbH, Model
FW7218M/24
Actuator - LA28/LA32
• Input voltage: 24V
• Noise level: ≤ 45 dB(A)
• IPX5
• Duty cycle: Max. 10% or 2 min. use then 18
min. without use
Handset - HB 30
• IPX6
Detachable parts
• Handset - HB 30
• Battery - BA 18
Applied part
• Handset - HB 30
Environmental conditions
• Operating temperature: +5°C to +40°C
• Storage temperature: -10°C to +50°C
• Relative humidity: 20% to 80%
• Atmospheric pressure: 700 to 1060 hPa
EN
LA28/LA32
Ambient and surface temperature
The table below illustrates how long time it
takes for the surfaces of the device to reach
operating temperature from -10°C and+50°C.
Surface temperature
-10°C
to
+50°C
to
52
HB30
CH01
BA18
Plug
+5°C
108 min
+40°C
129 min
etac.com