Coltene PerFect TCS II Manual De Usuario página 13

Ocultar thumbs Ver también para PerFect TCS II:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 46
Owner's Guide
1.2 EN/IEC 60601-1-2 Table 2
The Perfect
TCSII unit is intended for use in the electromagnetic environment specified below. The customer or the end user of the Perfect
®
TCSII unit should assure it is used only in such an environment.
Immunity Test
IEC60601 test level
Electromagnetic
+- 6kV contact
discharge (ESD)
+- 8kV air
IEC 61000-4-2
Electrical fast transient/
+-2kV for power supply lines
burst
+-1kV for input/output lines
IEC 61000-4-4
Surge
+-1kV differential mode (line-line)
IEC 61000-4-5
+-2kV common mode (line-earth)
Voltage dips, short
<5% UT (>95% dip in UT) for 0.5 cycle
interruptions and
voltage variations on
40% UT (60% dip in UT) for 5 cycles
power supply input
lines
70% UT (30% dip in UT) for 25 cycles
IEC 61000-4-11
<5% UT (>95% dip in UT) for 5 seconds
Power frequency
3A/m
(50/60Hz) magnetic
field
IEC 61000-4-8
Note UT is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
EN
Compliance Level
+- 6kV contact
+- 8kV air
+-2kV for power supply lines
+-1kV for input/output lines
+-1kV differential mode (line-line)
+-2kV common mode (line-earth)
<5% UT (>95% dip in UT) for 0.5 cycle
40% UT (60% dip in UT) for 5 cycles
70% UT (30% dip in UT) for 25 cycles
<5% UT (>95% dip in UT) for 5 seconds
3A/m
®
Intended Electromagnetic
Environment
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Mains power quality should
be that of a typical commer-
cial or hospital environment.
If the user of the Perfect
®
TCSII unit requires continued
operation during power mains
interruptions, it is recom-
mended that the Perfect
®
TCSII unit be powered from
an uninterruptible power sup-
ply with sufficient capacity to
run the unit for the maximum
required time of interruption.
Power frequency magnetic
fields should be at levels
characteristic of a typical loca-
tion in a typical commercial or
hospital environment.
13

Publicidad

Tabla de contenido
loading

Tabla de contenido