Display: True RMS
Accuracy guarantee: 5%~100% of range. Short circuit allows least significant digit ≤5.
Input impedance: ≥10 MΩ
Frequency response: 40 Hz~400 Hz
The AC crest factor will reach 3.0 at 3000 counts, while only ≤1.5 at 6000 counts. The additional
error should be added according to the crest factor of a non-sinusoidal wave:
a) Add 4% when crest factor is 1~2
b) Add 5% when crest factor is 2~2.5
c) Add 7% when crest factor is 2.5~3
For voltage frequency monitoring, the resolution is 0.1 Hz and accuracy is ±(0.1%+3). The input
amplitude should be ≥10% of range.
5. LPF ACV
Range
600.0 V
1000 V
Display: True RMS
Accuracy guarantee: 5%~100% of range. Short circuit allows least significant digit ≤5.
Input impedance: ≥10 MΩ
Frequency response: 40 Hz~200 Hz
The AC crest factor will reach 3.0 at 3000 counts, while only ≤1.5 at 6000 counts. The additional
error should be added according to the crest factor of a non-sinusoidal wave as follows:
a) Add 4% when crest factor is 1~2
b) Add 5% when crest factor is 2~2.5
c) Add 7% when crest factor is 2.5~3
The -3 dB frequency of LPF is about 2.5 kHz
Only manual range for LPF ACV. Use the RANGE button to change the range.
For voltage frequency monitoring, the resolution is 0.1 Hz and accuracy is ±(0.1%+3). The input
amplitude should be ≥10% of range.
6. LoZ V~
Range
600.0 V
1000 V
Display: True RMS
Accuracy guarantee: 5%~100% of range. Short circuit allows least significant digit ≤5.
Input impedance: About 2 kΩ
Frequency response: 40 Hz~400 Hz
The AC crest factor will reach 3.0 at 3000 counts, while only ≤1.5 at 6000 counts. The additional
error should be added according to the crest factor of a non-sinusoidal wave as follows:
a) Add 4% when crest factor is 1~2
b) Add 5% when crest factor is 2~2.5
c) Add 7% when crest factor is 2.5~3
When the measured voltage is above 220 V, the continuous measurement time cannot exceed 30 s
and the rest interval should be more than 30 s.
For voltage frequency monitoring, the resolution is 0.1 Hz and accuracy is ±(0.1%+3). The input
amplitude should be ≥10% of range.
Resolution
0.1 V
1 V
Resolution
0.1 V
1 V
Accuracy
Overload protection
± (2.0%+5)
Accuracy
Overload protection
± (2.0%+5)
GB
1000 A
1000 A
13