11. Capacitance (
)
Range
60.00 nF
600.0 nF
6.000 μF
60.00 μF
600.0 μF
6.000 mF
60.00 mF
•
Measurement result = displayed value – capacitance of open-circuit test leads
"REL" measurement mode is recommended for capacitance ≤1 μF
•
•
Accuracy guarantee: 5%~100% of range
12. Frequency/Duty ratio (Hz%)
Range
10 Hz~1 MHz
10.0%~90.0%
•
Frequency input amplitude:
10 Hz~100 kHz: 250 mVrms ≤ input amplitude ≤ 20 Vrms
100 kHz~1 MHz: 600 mVrms ≤ input amplitude ≤ 20 Vrms
•
Duty ratio:
10%~90%: for square waves of 10 Hz~1 kHz
30%~70%: for square waves of 1 kHz~10 kHz
2 Vpp ≤ input amplitude ≤20 Vpp
13. Non-contact AC voltage sensing (NCV)
Range
NCV
Resolution
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
0.001 mF
0.01 mF
Resolution
0.01 Hz~1 kHz
0.1%
Accuracy
To start sensing, move the NCV
sensor (upper tip) close to a wire.
If no voltage is sensed, the LCD
displays "EF". As the intensity of
the voltage detected increases,
m ore s e gm e nt s " — " w i l l b e
displayed, with buzzer and LED
flashing more frequently.
Accuracy
Overload protection
± (4.0%+25)
± (4.0%+5)
± (10.0%+9)
Accuracy
Overload protection
± (0.1%+3)
± (2.6%+7)
Overload protection
1000 V DC/AC
1000 A DC/AC
1000 V DC/AC
15
GB