Amprobe 5XP-A Manual De Uso página 23

Ocultar thumbs Ver también para 5XP-A:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Test current: 1.2 mA (approximate)
Accuracy: ±(1.5 % rdg + 3 dgts)
Open circuit volts: 3.0 dc typical
LOGIC TEST
Thresholds Logic 1 (Hi): 2.8 V ± 0.8 V
Thresholds Logic o (Lo): 0.8 V ± 0.5 V
Test voltage: TTL 5 VDC
OVERLOAD PROTECTION
Voltage, Resistance, Diode, Continuity, Logic:
600 VDC or 600 VAC rms
Current: 2 A / 1000 V fast blow ceramic fuse 6.3 × 32 mm
mA JACK: Input warning detects wrong switch / input jack
configuration
AUXILIARY FEATURES
DATA HOLD: Freeze the latest reading on the display.
RANGE: Execute manual range mode.
AUTO Power off: After auto power off, press (RANGE) button
to restart the meter, and the last reading of measurement
will be returned to the display.
35XP-A Electrical Specifications
(at 23°C ± 5°C, < 75 % R.H.)
DC VOLTS
Ranges: 400 mV, 4 V, 40 V, 400 V, 600 V
Accuracy: ± (0.5 % rdg + 1 dgt)
AC VOLTS (45 Hz to 500 Hz)
Ranges:400 mV, 4 V, 40 V, 400 V, 600 V
Accuracy: ± (1.5 % rdg + 5 dgts) 45 Hz to 100 Hz on 400 mV
range: ± (1.5 % rdg + 5 dgts) on 4 V to 600 V ranges
DC CURRENT
Ranges: 400 µA, 4000 µA, 40 mA, 400 mA, 2 A
Accuracy: ± (1.0 % rdg + 2 dgts) on 400 µA to 400 mA ranges:
± (2.0 % rdg + 3 dgts) on 2 A range
16
GlobalTestSupply
www.
.com
Find Quality Products Online at:
sales@GlobalTestSupply.com

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

15xp-b35xp-a

Tabla de contenido