Status of the foot switches
Symbol
MONO1
MONO2
BI
MONO2
MONO1
BIAUTO
Status of the neutral electrode
Symbol
42
Status
Foot switch 1
Foot switch 2
Foot switch 2
Foot switch 1
Foot switch 1
BIAUTO-
function
Status
HBS x-touch 300 argon
Meaning
-
MONOPOLAR1-
connections can be
controlled with foot
switch 1
-
MONOPOLAR2-
connections can be
controlled with foot
switch 2
-
MONOPOLAR2-
connections can be
controlled with foot
switch 2
-
BIPOLAR- connection
can be controlled with
foot switch 1
-
MONOPOLAR1-
connections can be
controlled with foot
switch 1
-
BIAUTO- function
active. This allows an
automatic start with
recognition of tissue in
the bipolar application
Meaning
Non divided neutral
electrode is detected
Divided neutral electrode is
detected
Neutral electrode not
detected or not connected