EN
8.13 Operating errors
In the case of an operating error, the following display appears:
The following error messages can occur:
Error - display
Cause of error
Neutral electrode not connected,
NE Error - No 1
not known, or NE is defective
Mono1 - Cut hand/
Cut foot switch or button on handle
FS1 Error
in terminal Mono 1
Mono1 - Coag
Coag foot switch or button on
hand/ FS1 Error
handle in terminal Mono 1
Mono2 - Cut hand/
Cut foot switch or button on handle
FS2 Error
in terminal Mono 2
Mono2 - Coag
Coag foot switch or button on
hand/ FS2 Error
handle in terminal Mono 2
The pressure in the cylinders is below
the required minimum or there is an
error in the pressure measurement
system
"ARGON level LOW"
The flow volume set at the display
"ARGON no FLOW"
cannot be reached
If one or more numerical codes are displayed, or if the device is indicating any other recognizable
defects, please refer to service@hebumedical.de.
◼
HF current does not flow when the foot switch is actuated. The system only reacts with a
tone signal and the display changes colour.
Please check:
➔
That all terminals have been correctly connected
➔
The settings of the foot switch
➔
Whether the neutral electrode has been recognized
◼
The neutral electrode is not or not correctly detected
Check all terminals and touch the field with the symbol of the neutral electrode for detection of the
NE. If the neutral electrode is still not recognized, exchange it where appropriate.
◼
It is not possible to switch the machine on or off.
Check whether the power cable has been correctly connected and that the socket is intact. If the
error persists, inform the specialist personnel.
◼
The system works perfectly, but the settings cannot be saved.
Inform the specialist personnel.
56
Remedying errors
Check the terminal, touch the NE symbol
on the touchscreen for detection, where
applicable exchange NE and/or
connecting cable.
Check the terminals, check the settings,
exchange accessories if necessary. If
error persists, inform specialist personnel.
Check terminals and settings, exchange
accessories if applicable, if error persists,
exchange argon module
Check the terminals and gas cylinder. If the
error persists, exchange argon module
HBS x-touch 300 argon