HBS x-touch 300 argon
6 Specifications
Mains connection
Supply voltage
Nominal frequency
Output power in monopolar operation
Pure cut
Blend cut
Polypectomy mode
Cutting in water environment (TUR)
Soft coagulation
Forced coagulation
Spray coagulation
Output power in bipolar operation
Blend Cut
Coagulation
Cutting in water environment (TUR)
AUTO STOP
Argon
Argon purity
Max. admission pressure
Argon gas cylinders
Regulation
Safety measures
Type
Protection class
Equipotential connection
Safety
Electromagnetic compatibility
Safety of high-frequency surgical units
100-260 Volt, 50 - 60 Hz.
500 kHz
300 W at 500 Ω
250 W at 500 Ω
100 W (peak: 250W) at 250 Ω
300 W at 500 Ω
150 W at 250 Ω
150 W at 500 Ω
80 W at 1500 Ω
100 W at 100 Ω
100 W at 100 Ω
120 W at 100 Ω
Yes
mind. 4.6 (99.998 %)
2.5 bar
1 or 2 cylinders with 5 L
0.3 – 6.0 L/min in 0.1 L/min steps
CF
I
Yes
In accordance with DIN EN ISO 60601-1
In accordance with DIN EN ISO 60601-1-2
In accordance with DIN EN ISO 60601-2-2
EN