Descargar Imprimir esta página

HEBU medical HF 9514 Instrucciones Para El Uso página 50

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

EN
8.5 Foot switch settings
After connecting the foot switch at the back, the required mode must be set at the touchscreen. To set
the mode, touch the field indicating the foot switch symbol. This changes the text which appears
above the symbol.
It is important to ensure that the instruments are connected to the connectors, which are selected by
footswitch. For example: In Monopolar 1 connected instruments can only be operated with the
settings MONO1/MONO2 or MONO1 + BIAUTO of the foot switch.
8.6 Handles with buttons
Activating the handle with buttons takes place independently of the foot switch settings. These cannot be
activated using the foot switch.
8.7 Neutral electrode settings and attachment
In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognize the neutral
electrode automatically (cf. illustration on the left and chapter 5.2 "
operating modes and settings
with divided and undivided contact surfaces.
If the displayed neutral electrode type does not agree with the used
neutral electrode on the patient, a defective neutral electrode or a
Please note the following information when attaching the
50
").The HBS x-touch 300 argon recognizes neutral electrodes
neutral electrode type, this is indicative of poor adhesion of the
defective cable!
neutral electrode
HBS x-touch 300 argon
Overview of symbols for

Publicidad

loading

Este manual también es adecuado para:

Hbs x-touch 300 argon