HBS x-touch 300 argon
8.2 Connection of argon
The HBS x-touch 300 argon is able to work both with gas from a cylinder or from a gas transmission
system.
Gas flow from gas transmission system
•
Close the hose at one of the two connection points (D or G).
Use of gas cylinders
•
Secure the gas cylinder (A) with the provided fixture (B)
•
Do NOT remove the identification label
•
Mount an original HEBU pressure relief valve HEBU (C) at the
gas cylinder (article number: HF 9530-02)
•
Connect the hose (D) of gas cylinder 1 to
connection point 1 (25) and the gas pressure
measurement line (E) from the valve to terminal
26. Connect the hose (G) of gas cylinder 2 to
connection point 2 (27) and the gas pressure
measurement line 2 (F) to the corresponding
terminal 28.
the HBS X-touch 300 argon automatically switches to the full gas
Use ultra-pure argon gas with a purity level of 99.996 % or
higher. Using gas with a lower purity can place the safety of
When the pressure in the active gas cylinder drops to below 5%
cylinder. If both cylinders are empty, the message "Argon level
LOW" appears on activation at the display.
patients in jeopardy.
EN
47