FreeStyle Comfort
5. Turn on the FreeStyle Comfort by pressing the power
button and select your prescribed flow setting utilizing
the increase (+) or decrease (-)flow buttons.
A brief audible alarm will sound to indicate the device
has powered on.
Note: When the FreeStyle Comfort is initially turned on the
default flow setting is 2. If the setting is changed, the unit
will power on to the previous flow setting.
6. Position the nasal cannula on your face per the
manufacturer's instructions provided with the cannula
and breathe normally. You should feel a puff of oxygen
delivered through the nasal cannula each time you
inhale.
7. To change the flow setting, press the increase (+) or
decrease (-) flow buttons to the desired flow rate. It is
normal to hear a difference in sound as you change the
flow settings.
12 - ENG
PN MN250-C4 H | User Manual
Pulse Volume (mL)*
Setting
Breath
1
2
Rate
15
14
28
20
10.5
21
25
8.4
16.8
30
7
14
35
6
12
40
5.3
10.5
*Tolerance is +/- 15%
8. Turn the FreeStyle Comfort off by pressing the power
button.
Note: The display screen will enter standby mode and dim
after 20 seconds. Press any button to wake the screen out
of standby mode.
Special product features
UltraSense
®
- This sensitive pulse dose trigger detects the
pressure of breath, and releases the pulse of oxygen you
need, when it is most critical.
autoDOSE
- Should a no breath alarm occur, the Free-
™
Style Comfort will pulse oxygen at its current flow setting
at 20 breaths per minute to its user until the alarm state is
resolved.
3
4
5
42
56
70
31.5
42
52.5
25.2
33.6
42
21
28
35
18
24
30
15.8
21
26.3