Storz IMAGE1 S TC 200 Manual De Instrucciones página 70

Ocultar thumbs Ver también para IMAGE1 S TC 200:
Tabla de contenido

Publicidad

62
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
The IMAGE1 S™ is intended for use in the electromagnetic environment specified below.
The user of IMAGE1 S™ should make sure that it is used in such an environment.
Immunity tests
EN/IEC 60601 test level
Electrostatic discharge
± 6 kV Contact discharge
(ESD) acc. to
± 8 kV Air discharge
IEC 61000-4-2
Rapid transient electrical
± 2 kV for power lines
interferences/bursts acc.
± 1 kV for input and
to IEC 61000-4-4
output lines
Surges
± 1 kV line(s) to line(s)
acc. to IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short inter-
ruptions and fluctuations
<5 % U
*
T
of the supply voltage
(>95 % dip of the U
acc. to IEC 61000-4-11
for ½ cycle
40 % U
T
(60 % dip of the U
for 5 cycles
70 % U
T
(30 % dip of the U
for 25 cycles
<5 % U
T
(>95 % dip of the U
for 5 seconds
Magnetic field at the
supply frequency
3 A/m
(50/60 Hz)
acc. to IEC 61000-4-8
* Note: U
is the mains alternating voltage before applying the test levels.
T
Information on
electromagnetic
compatibility (EMC)
Table 2
Compliance level
± 6 kV Contact discharge
± 8 kV Air discharge
± 2 kV/1 kV
± 1 kV for input and
output lines
± 1 kV differential mode
± 2 kV common mode
<5 % U
*
T
)
(>95 % dip of the U
)
T
T
for ½ cycle
40 % U
T
)
(60 % dip of the U
)
T
T
for 5 cycles
70 % U
T
)
(30 % dip of the U
)
T
T
for 25 cycles
<5 % U
T
)
(>95 % dip of the U
)
T
T
for 5 seconds
3 A/m
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30 %.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
If the user of the IMAGE1 S™ require continued operation
during power mains interruptions, it is recommended that the
IMAGE1 S™ be powered from an uninterruptible power supply
or a battery.
If image distortion occurs, it may be necessary to position the
IMAGE1 S™ further from sources of power frequency magnetic
fields or to install magnetic shielding. The power frequency
magnetic field should be measured in the intended installation
location to assure that it is sufficiently low.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Image1 s tc 300Image1 s tc 301Image1 s tc 302Image1 s tc 304

Tabla de contenido