Stacking
Accessories required:
Models:
Stock no.:
6060
50027660
6120
6200
50027661 50027661
Incubator models 6060, 6120 and 6200 can be stacked as
follows:
A max. of 2 on top of each other and 2 stacks side-by-side;
clearance between stacks 25 mm.
The procedure is carried out in the following manner:
Remove the plastic rivets.
Put down the facing.
Secure the stacking elements and angular facing with head
screws.
Place the second incubator on top of the first one. The
stacking elements fit into the hollow feet of the upper
incubator to secure it in place.
XJ
XT
i
I
50012338
- 7 -