EN-20
ADVENTURER BALANCES
Confirm the default values displayed (Water Temp)
are correct.
To edit the default values, touch Item Settings.
The Settings screen appears.
The balance calculates water density based on the
water temperature value entered (look-up table).
Measure the actual water temperature using a
precision thermometer.
To adjust the Water Temperature or Oil Density
values,
touch the Water Temp or Oil Density button.
Numeric input windows appear.
Key in the desired value,
then press Save.
The display returns to the previous
screen with the new value
highlighted.
To return to the Density
Determination home screen, touch
Exit.
Press Begin Density Calculation.
Step 1 of 3
Weigh the Un-Oiled Sample in Air.
Follow the screen instructions then press Accept to store the
dry sample weight (in air).